Ceruloplasmln assay, Alinity c Ceruloplasmin Reagent Kit (2019-06-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lots.B. Alinity c Ceruloplasmin Reagent Kit
Manufacturer
Sentinel Ch Spa
Via Robert Koch, 2
Milan
20152
ITALY
09P9320
All serial numbers.A. Ceruloplasmln assay
Manufacturer
Sentinel Ch Spa
Via Robert Koch, 2
Milan
20152
ITALY
6K91-30

Introduction

Affected Products

A. Ceruloplasmln assay
B. Alinity c Ceruloplasmin Reagent Kit

Reason

Package insert of the ceruloplasmin assay lists the EDTA as suitable anticoagulant for sampling collection tube. However, the EDTA sample tube storage claims are not met with the new studies performed. In particular, the stability claims of 2 weeks at 2-8 degrees celsius and 8 days at room temperature are not confirmed, both for dipotassium and for tripotassium EDTA.