Health Products

Elekta UNITY (2019-06-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Elekta Limited
Linac House, Fleming Way
Crawley, Sussex
RH10 9RR
UNITED KINGDOM
Elekta  UNITY1536549

Introduction

Affected Products

Elekta UNITY

Reason

If the Set To Work (STW) phase 2 of the installation is not conducted in the correct order, the MR to MV offset values cannot be exported to the EQ database at STW phase 2 and (0,0,0) values are set instead by software installer in order to run clinical workflow tests.