Health Products

Tina-Quant Myoglobin Gen.2 (2019-06-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
34986001 and higher.
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
Tina-Quant Myoglobin Gen.24580010190

Introduction

Affected Products

Tina-Quant Myoglobin Gen.2

Reason

There is quality controls (QCS) imprecision and calibration failures with the Tina-Quant Myoglobin Gen.2 (Myo2) assay on COBAS c 311/501/502 with reagent lot #34986001. COBAS INTEGRA400 plus/COBAS c 701-702 are not impacted.