AQUIOS IMMUNO-TROL Low Cells (2019-07-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberTelephoneBrand nameCompaniesFaxModel or catalog number
6180091 (6180091K)+353 (0) 65 683 1100
AQUIOS IMMUNO-TROL Low Cells
Manufacturer
Beckman Coulter Diagnostics Limited
Lismeehan
+353 (0) 65 683 1122 B25700

Introduction

Affected Products

AQUIOS IMMUNO-TROL Low Cells

Reason

Some Beckman Coulter customers have reported quality control (QC) failure on the AQUIOS system when performing testing with the affected lot. Beckman Coulter has confirmed through internal testing of this product lot that the system may include a significant portion of the monocyte population in the lymphocyte region during automated data analysis due to insufficient separation of the two populations in the side scatter vs. CD45 plot.