Brilliance 40-64 Slice CT System, Brilliance ICT System - System, Brilliance ICT SP System - Main, Ingenuity CT System (2019-06-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
D. Ingenuity CT System728323
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
B. Brilliance ICT System- System728306
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
C. Brilliance ICT SP System - Main728311
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
A. Brilliance 40-64 Slice CT System728231

Introduction

Affected Products

A. Brilliance 40-64 Slice CT System
B. Brilliance ICT System- System
C. Brilliance ICT SP System - Main
D. Ingenuity CT System

Reason

There are 3 scenarios users may experience which result in identical images that are incorrectly labeled as phase tolerance images:
  1. When performing a Step and Shoot acquisition with phase tolerance selected, images may be reconstructed identically (a single phase), but are labeled as different phases. This occurs when ""Start Final Recon"" is selected prior to display of the ECG wave at the top of the acquisition window.
  2. When performing a Step and Shoot acquisition with phase tolerance selected, images may be reconstructed identically (a single phase), but are labeled as different phases. This occurs if ECG leads become disconnected mid-acquisition or when the acquisition is halted prematurely due to an application crash.
  3. By system design, in certain instances when a patient exhibits an arrhythmia or varying heart rate during a Step and Shoot acquisition, images from the acquisition following the heart rate variation, may be labeled in the incorrect phase.