IMMULITE, 1000 and 2000 System- Progesterone Assay
Reason
Siemens Healthcare Diagnostics INC. Has determined that there is a potential for low discordant progesterone results on a subset of patient samples. Our investigation of customer complaints suggests the presence of a potential interferent with the assay. Based on available data, it is estimated the occurrence is <1%. Due to the nature of initial complaints received, the investigation to date has focused on in vitro fertilization (IVF) patient sample performance.