A. DIMENSION CLINICAL CHEMISTRY SYS. - FERRITIN (FERR) FLEX REAGENT CARTRIDGE
B. DIMENSION XPAND SYSTEM - FERRITIN (FERR) ASSAY
C. DIMENSION EXL SYSTEM - FERRITIN (FERR) ASSAY
Reason
Siemens Healthcare Diagnostics has confirmed that the Dimension® Ferritin (FERR) Flex® reagent lot listed in Table 1 may exhibit a negative bias during the open well stability time period. Siemens has determined that FERR values can potentially decrease by -50% at a concentration of 22 ng/mL [μg/L] and -48% at a concentration of 285 ng/mL [μg/L] after a 10 day open well period. Siemens has also determined that FERR values can potentially decrease by approximately -15% after a 3 day open well period. Quality control and patient samples are affected similarly.