BIOGRAPH MCT SYSTEM (2019-07-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
11069B. BIOGRAPH MCT-S WITH TRUE V (40)
Manufacturer
Siemens Medical Solutions USA Inc.
Molecular Imaging Group,
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES
10248671
11002C. BIOGRAPH MCT FLOW-4R
Manufacturer
Siemens Medical Solutions USA Inc.
Molecular Imaging Group,
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES
10529159
11044A. BIOGRAPH MCT-S (40)
Manufacturer
Siemens Medical Solutions USA Inc.
Molecular Imaging Group,
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES
10248668

Introduction

Affected Products

A. BIOGRAPH MCT-S (40)
B. BIOGRAPH MCT-S WITH TRUE V (40)
C. BIOGRAPH MCT FLOW-4R

Reason

Biograph MCT or MCT Flow systems running petsyng VG70A software may have issues where the lSO QC (quality check) does not execute correctly on the newest system designs. Invalid time tags may result in PET scan aborts and failure of PET image reconstruction. These issues may result in the necessity for patient rescan.