Health Products

ASSAYED CHEMISTRY CONTROL PREMIUM PLUS (2019-07-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
961UE
Manufacturer
Randox Laboratories Ltd.
55 Diamond Road
Crumlin, County Antrim
BT29 4QY
UNITED KINGDOM
ASSAYED CHEMISTRY CONTROL PREMIUM PLUSCAL2351

Introduction

Affected Products

ASSAYED CHEMISTRY CONTROL PREMIUM PLUS

Reason

Randox have realigned the RX Analyser Series calibration targets by 10% for AST and 9% for ALT in line with the Mean of All Instrument method target from calibrator lot 961UE onwards. No other instruments are affected. The target and range for the associated Quality Control Material has been revised to align with this change. Updated RX control targets are available at randox.com.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime