Health Products

Disposable Hyperinflation System (2019-07-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Ambu A/S
Baltorpbakken 13
Ballerup
2750
DENMARK
Disposable Hyperinflation System900000604

Introduction

Affected Products

Disposable Hyperinflation System

Reason

Hyperinflation system has been packaged with the manometer, which did not include the shelf life of the manometer. The root cause was determined to be that the shelf life of the manometer was moved from ifu to pouch labeling. During assembly, manometer is opened from the pouch and attached to the hyperinflation bag. However, accompanying packaging (plastic pouch) for the manometer is discarded. This results in a loss of shelf life information for the finished good.