Spectra Optia Apheresis System (2019-07-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1P02482
Manufacturer
Terumo BCT Inc.
10811 W. Collins Avenue
Lakewood
80215
Colorado
UNITED STATES
Spectra Optia Apheresis System61000

Introduction

Affected Products

Spectra Optia Apheresis System

Reason

On 2019-05-16 complaints entered into Terumo's database identified a trend of reports from the field stating that an alarm rate (T0AirEmbolism) in Optia Gen 2 e-boxes for when the low-level sensor sees air and the safety computer sees the return pump still rotating in the direction of patient were being  triggered without cause. Root cause investigation shows the issue being related to the current pump driver CCA (777200-544) that was implemented with Gen 2 e-box.