On 2019-05-16 complaints entered into Terumo's database identified a trend of reports from the field stating that an alarm rate (T0AirEmbolism) in Optia Gen 2 e-boxes for when the low-level sensor sees air and the safety computer sees the return pump still rotating in the direction of patient were being triggered without cause. Root cause investigation shows the issue being related to the current pump driver CCA (777200-544) that was implemented with Gen 2 e-box.