ATELLICA CH 930 ANALYZER (2019-08-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
CM00237
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
ATELLICA CH 930 ANALYZER11067000

Introduction

Affected Products

ATELLICA CH 930 ANALYZER
 

Reason

Siemens Healthcare Diagnostics, Inc. is informing customers that the Atellica CH 930 Analyzer may report incorrect serum indices for Hemolysis (H) and Lipemia (L) when using the UN_c test as a donor method, and the Hemolysis Icterus Lipemia (HIL) function is enabled. Due to bubbles in the reaction cuvette, there is a potential for falsely depressed H results (e.g. H index of 3 or higher, reported as ¿1 ) at a rate of 0.3-1% and falsely elevated L results (e.g. L index of 0 reported as ¿ 1) at a rate of 2.9%. In all cases of falsely depressed H results, the L results will be elevated ¿1. For samples not using UN_c as the donor method the indices are not affected. THE DONOR METHOD THE INDICES ARE NOT AFFECTED.