Health Products

AutoTac Handle (2019-07-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Biohorizons Implant Systems Inc.
2300 Riverchase Center
Birmingham
35244
Alabama
UNITED STATES
AutoTac Handle400-200

Introduction

Affected Products

AutoTac Handle

Reason

Biohorizons determined certain lots of the AutoTac Handle of 400-200 instrument were assembled with a non-conforming reduced size spring spacer. The non-conforming spring spacer of .049 Inches was assembled in the handle instead of the conforming .230 Inches spring spacer.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime