ADVIA Centaur System - Folate Assay (2019-07-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Products Limited
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
ADVIA Centaur System - Folate Assay6367974

Introduction

Affected Products

ADVIA Centaur System - Folate Assay

Reason

Siemens Healthcare Diagnostics has confirmed the potential for the ADVIA Centaur homocysteine assay (100 test kit smn 10310374; 500 test kit smn 10310375) to intermittently interfere with the folate assay when testing is performed on the ADVIA Centaur, ADVIA Centaur XP or ADVIA Centaur XPT systems. The interference is observed intermittently with only the first replicate of folate when the assay is tested immediately following homocysteine assay testing.