Health Products

Seca 402 & 403 (2019-07-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Seca GMBH & co. KG
HAMMER STEINDAMM 3-25
Hamberg
22089
GERMANY
B. Seca 4034030000009
6002142000806
Manufacturer
Seca GMBH & co. KG
HAMMER STEINDAMM 3-25
Hamberg
22089
GERMANY
A. Seca 4024020000009

Introduction

Affected Products

A. Seca 402
B. Seca 403

Reason

The Seca 402 and Seca 403 baby scale carts are equipped with wheels.  Unanticipated loosening of the steering fork that attaches the wheels to the cart can occur. In the worst case, the steering fork and wheel can detach from the cart.