Medtronic has identified that flushing the delivery system after the tether has been cut may cause the free end of the tether to enter the ventricular chamber and tangle. If this happens, the tangle could potentially prevent tether removal from the delivery system and potentially prevent release of the Micra pacemaker. Therefore, medtronic is updating the IFU and implant procedure tip card to include the recommendation below:
once the tether is cut, do not flush the delivery system until the tether has been fully removed.