Anti Human Anti Hemophilus Influenza EIA Kit (2019-07-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
413170
Manufacturer
The Binding Site Group Limited
8 Calthorpe Road
Birmingham
B15 1QT
UNITED KINGDOM
Anti Human Anti Hemophilus Influenza EIA KitMK016

Introduction

Affected Products

Anti Human Anti Hemophilus Influenza EIA Kit

Reason

Investigations into customer complaints have shown an average positive bias of 27.3% With a probability of low QC failure of 42% using manual testing methods. A positive bias of 43.7% With a probability of failure of 45% (low QC) and 43% (high QC) was observed when using automated testing methods. It was identified that the positive bias is more pronounced at lower concentrations, and presents inconsistently across individual plates.