Zimmer Biomet is conducting a medical device recall for the Vanguard XP-XP Tibial Tray due to the possibility of the locking bar not fully engaging. The issue was identified through a complaint investigation that determined that the edge of the tray may contain a burr that can prevent the locking bar from fully engaging. Burrs are not present on all trays and the surgical technique indicates that there should be user confirmation of proper engagement through visual (the locking bar arm, clasp, and polyethylene bearing must be flat on the baseplate in all directions) as well as auditory (clicks in place) checks. Based on complaint reports, when engagement was not initially fully achieved, the visual/audible check detected this condition intra-operatively. Following detection, use of the insertion tool and rigid instrumentation allowed for full engagement.