Affected Products
CentriMag Extracorporeal Blood Pumping System - Motor
Abbott is advising our customers that we have received reports of CentriMag systems experiencing motor and pump issues resulting from electromagnetic interference (EMI). Due to this EMI exposure, customers have experienced unexpected console screen blanking, low flow or no flow, and motor behavior including noise, vibration and heating have been reported at a rate of 0.45%. Investigation has identified CentriMag motors with certain serial numbers between l05333-0001 and l06608-0024 with an increased susceptibility to EMI In a small number of cases, this issue has been associated with serious injuries due to hemodynamic compromise resulting from interruption of support. This issue can be mitigated through recalibration of the motor.