Health Products

CentriMag Extracorporeal Blood Pumping System - Motor (2019-08-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
SN's between L05333-0001 and L06608-0024
Manufacturer
Thoratec Switzerland GmbH
Technoparkstrasse 1
Zürich
8005
SWITZERLAND
CentriMag Extracorporeal Blood Pumping System - Motor201-10002

Introduction

Affected Products

CentriMag Extracorporeal Blood Pumping System - Motor

Reason

Abbott is advising our customers that we have received reports of CentriMag systems experiencing motor and pump issues resulting from electromagnetic interference (EMI). Due to this EMI exposure, customers have experienced unexpected console screen blanking, low flow or no flow, and motor behavior including noise, vibration and heating have been reported at a rate of 0.45%. Investigation has identified CentriMag motors with certain serial numbers between l05333-0001 and l06608-0024 with an increased susceptibility to EMI In a small number of cases, this issue has been associated with serious injuries due to hemodynamic compromise resulting from interruption of support. This issue can be mitigated through recalibration of the motor.