Health Products

COBAS OMNI LYSIS REAGENT (2019-08-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
F00997COBAS OMNI LYSIS REAGENT
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
6997538190

Introduction

Affected Products

COBAS OMNI LYSIS REAGENT

Reason

Following the information provided by our manufacturer on internal investigation of customer complaints for the cobas omni Lysis reagent  (cat # 06997538190, lot f00997), we would like to inform you of a situation describing increase rate of unexpected reactive results with the cobas West Nile Virus assay  when tested on the cobas 6800-8800 analyzers. The manufacturer advised us of this on 14-Aug-2019.