Health Products

HFA Review (accessory of Humphrey Field Analyzer 3 (HFA3)) (2019-09-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable.
Manufacturer
Carl Zeiss Meditec, Inc.
5160 Hacienda Drive
Dublin
94568
California
UNITED STATES
HFA Review (accessory of Humphrey Field Analyzer 3 (HFA3))HFA 830

Introduction

Affected Products

HFA Review (accessory of Humphrey Field Analyzer 3 (HFA3))

Reason

HFA review software version 1.0.1 which is compatible with HFA 830, 840, 850 and 860 models can possibly cause loss of test results/medical record information leading to a delay in treatment. There is a potential for a delay of treatment that could result in minor deterioration in a patient's vision (loss of less than a line on a visual acuity chart). The HFA review software is an accessory to the HFA instrument and the HFA instrument is not affected.