Medtronic is voluntarily retrieving specific Synchromed II Implantable Drug Infusion Pumps after investigating complaints related to permanent motor stall. This voluntary recall is being conducted due to the potential for the presence of a foreign particle inside the pump motor assembly which could interfere with motor gear rotation and lead to a permanent motor stall. The source of the foreign particle has been identified and eliminated. A permanent pump motor stall will result in cessation of drug infusion therapy which may result in return of underlying symptoms and/or withdrawal symptoms.