Health Products

SynchroMed II - Programmable Pump (2019-10-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
SynchroMed II - Programmable Pump8637-20

Introduction

Affected Products

SynchroMed II - Programmable Pump

Reason

Medtronic is voluntarily retrieving specific Synchromed II Implantable Drug Infusion Pumps after investigating complaints related to permanent motor stall. This voluntary recall is being conducted due to the potential for the presence of a foreign particle inside the pump motor assembly which could interfere with motor gear rotation and lead to a permanent motor stall. The source of the foreign particle has been identified and eliminated. A permanent pump motor stall will result in cessation of drug infusion therapy which may result in return of underlying symptoms and/or withdrawal symptoms.