Steris AMSCO Eagle 3000 Prevacuum Sealer (2019-10-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Steris Mexico
790 Avante Ave., Industrial Park
Guadalupe, Nuevo Leon
67190
MEXICO
Steris AMSCO Eagle 3000 Prevacuum SealerAX00-XXX

Introduction

Affected Products

Steris AMSCO Eagle 3000 Prevacuum Sealer

Reason

Steris has identified that a small number of preventive maintenance kits for the Eagle 3000 small sterilizer were assembled with a 100 psig safety valve instead of a 40 psig valve, potentially resulting in the 100 psig valves being installed on sterilizers in the field. Use of the incorrect valve could result in the improbable scenario of water and steam leaking from the unit's fittings, or accumulation of steam in the chamber. Steris has not received any reports of adverse events related to the issue described above.