Health Products

Star Total Ankle Replacement - Sliding Core (2019-10-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Stryker GMBH
Bohnackerweg 1,
Selzach, Solothurn
2545
SWITZERLAND
Star Total Ankle Replacement - Sliding Core400-141

Introduction

Affected Products

Star Total Ankle Replacement - Sliding Core

Reason

Stryker has become aware of data indicating that patients implanted with the Star Total Ankle Replacement distributed prior to August 1, 2014 may experience a higher than expected risk of polyethylene fracture due to potential increase in polyethylene oxidation prior to or after implantation and potentially the implant geometry (as reported in published literature).