Health Products

Apollo - Main Unit (2019-10-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
s/n 08070501
Manufacturer
Villa Sistemi Medicali S.P.A.
via Delle Azalee 3
Buccinasco
20090
ITALY
Apollo - Main Unit9784120231

Introduction

Affected Products

Apollo - Main Unit

Reason

Following information received from field, it was found that the number of tilting cycles for some units used in high-workload environments had to be increased with respect to the initial specifications. After laboratory testing of the current angulation chain, it was found that it is very close to the design limit for metal fatigue and, under severe workload conditions might fail before the expected lifetime of 10 years.