Health Products

Ventana HE 600 System (2019-10-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer. Ventana HE 600 System
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
6917259001

Introduction

Affected Products

Ventana HE 600 System

Reason

Situation related to potential burnt coverslip module on the Ventana HE600 system (Cat # 06917259001). A leak in the staining module caused an electrical short and damage to the electrical component.