ACUSON SC2000 (2019-10-31)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
VB20
Manufacturer
Siemens Medical Solutions USA Inc.
685 East Middlefield Road
Mountain View
94043
California
UNITED STATES
ACUSON SC200010433816

Introduction

Affected Products

ACUSON SC2000

Reason

Siemens is notifying customers of a variation in the results associated with the eSie Flow SW license on the ACUSON SC2000 ultrasound system. The eSie Flow quantification results are calculated assuming a 0° Angle for all angles of acquisition, however, in software versions VB20 or VB21 the eSie Flow does not correct for the angle of the blood flow. There is an improbable probability that stroke volume, regurgitant volume and regurgitant fraction results will be estimated incorrectly under certain conditions, which may lead to potential patient misdiagnosis