During internal reagent testing it was shown that the linearity claim of the 1000umol/ml for the cobas Integra Fructosamine used on cobas Integra 400 Plus Analyzer could not be achieved anymore. As a consequence, the manufacturer reduced the upper measuring range to 900umol/l. The manufacturer advised us of this issue on 2019/10/28.