Bellavista 1000, Bellavista 1000(Neo) & Bellavista 1000E (2019-10-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
160012
Manufacturer
IMTMEDICAL AG
GEWERBESTRASSE 8
BUCHS
SWITZERLAND
B. Bellavista 1000 (Neo)301.100.060
170031
Manufacturer
IMTMEDICAL AG
GEWERBESTRASSE 8
BUCHS
SWITZERLAND
C. Bellavista 1000E301.100.100
More than 10 numbers, contact manufacturer.
Manufacturer
IMTMEDICAL AG
GEWERBESTRASSE 8
BUCHS
SWITZERLAND
A. Bellavista 1000301.100.000

Introduction

Affected Products
  1. Bellavista 1000
  2. Bellavista 1000 (Neo)
  3. Bellavista 1000E

Reason

Bellavista 1000 ventilators hardware generations G2/3/4/5/6 can experience the following ""intermittent failures during ventilation"": lack of acoustic high priority alarm, presence of a 'no alarm' condition, presence of a 'fail safe state'.
  • Lack of acoustic high priority alarm (continuous alarm tone) under specific conditions which may cause a delay in immediate action required to avert a life-threating situation.
  • Presence of a 'no alarm' condition during a disconnect under specific use conditions which may cause a system leakage and potential for loss in ventilation therapy without activation of alarm.
  • Presence of a 'fail safe state' under specific use conditions which may cause a device (ventilator) response by suspending ventilation to the patient.