Health Products

Flexitron Brachytherapy Remote Afterloading System (2019-11-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer. 
Manufacturer
Nucletron B.V.
Waardgelder 1
Veenedaal
3905 TH
NETHERLANDS
Flexitron Brachytherapy Remote Afterloading System136149A12-06

Introduction

Affected Products

Flexitron Brachytherapy Remote Afterloading System

Reason

This is concerning an afterloader system in which the internal check cable guide mechanism failed. Due to an obstruction during the check cable out drive, the internal guiding tube became displaced and the check cable became damaged.