Health Products

CIC Pro Clinical Information Center, ApexPro Telemetry Systems & CARESCAPE Central Station (2019-11-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
GE Medical Systems Information Technologies, Inc.
8200 W Tower Ave
Milwaukee
53223
Wisconsin
UNITED STATES
B. ApexPro Telemetry SystemsCARESCAPE TELEMETRY SERVER
More than 10 numbers, contact manufacturer.
Manufacturer
GE Medical Systems Information Technologies, Inc.
8200 W Tower Ave
Milwaukee
53223
Wisconsin
UNITED STATES
C. CARESCAPE Central Station2037318-007
More than 10 numbers, contact manufacturer.
Manufacturer
GE Medical Systems Information Technologies, Inc.
8200 W Tower Ave
Milwaukee
53223
Wisconsin
UNITED STATES
A. CIC Pro Clinical Information CenterCIC-XXXX-XXXXX-XXXXX

Introduction

Affected Products

A. CIC Pro Clinical Information Center
B. ApexPro Telemetry Systems
C. CARESCAPE Central Station

Reason

When connected to the Mission Critical and/or Information Exchange networks, certain versions of the CARESCAPE Telemetry Server, Apex Telemetry Server, CARESCAPE Central Station version 1 and Central Information Center systems were identified to have vulnerabilities to cyber-attack, which could result in a loss of monitoring and/or alarms during patient monitoring.