Health Products

Trace Elements Normal Range Serum Control (2019-11-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
C2540
Manufacturer
UTAK LABORATORIES INC.
25020 AVENUE TIBBITTS
VALENCIA
91355
California
UNITED STATES
Trace Elements Normal Range Serum Control66816

Introduction

Affected Products

Trace Elements Normal Range Serum Control

Reason

Incorrect value on the product IFU for the analyte of Manganese. The value was mistyped as a value of 2.69 and range of 2.29 - 3.09 on the IFU, when it was reported as 0.48 and 0.41 - 0.55 respectively.