Health Products

Stellar 100 and Stellar 150 (2019-12-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
20160123307 to 22171057208
Manufacturer
Resmed Limited
1 Elizabeth Macarthur Drive
Bella Vista
2153
AUSTRALIA
Stellar 100 and Stellar 15024140

Introduction

Affected Products

Stellar 100 and Stellar 150

Reason

Resmed received a confirmed case of an alarm buzzer not working under the following circumstances:  the device has a failed electronic component and, the device is stored without AC power connected for more than 36 hours leading to full depletion of the battery and, the device powers on automatically when connected to AC power without pressing the power switch.  The identified issue is a combination of software and a component failure that may occur in a small number of devices.  If a device with is issue is used in the condition described above, the audible alarms may not operate properly.