Health Products

DIGORA Optime, DIGORA Optime UV (2020-01-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 codes, contact manufacturer.
Manufacturer
Soredex Palodex Group OY
Nahkelantie 160
Tuusula
04300
FINLAND
DIGORA Optime, DIGORA Optime UVDXR 50

Introduction

Affected Products

DIGORA Optime, DIGORA Optime UV

Reason

Following customer complaints, Palodex identified an issue with power supply units supplied with our imaging plate systems in 2012-2013. Slow degradation of plastic insulation in the power supply unit DC plug's insulation may result in short-circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the power supply unit to fail. No fire or flames have been reported, nor harm to patients or users. We confirmed devices were distributed in Canada on December 10, 2019.