Health Products

AZURION 7 M20, B12 and B20 (2019-12-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable. 
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
B. AZURION 7 B12722 067
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
C. AZURION 7 B20722 068
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
A. AZURION 7 M20722 079

Introduction

Affected Products

A. AZURION 7 M20
B. AZURION 7 B12
C. AZURION 7 B20

Reason

Due to a software defect, when a digital subtracting angiography (DSA) is exported to the vessel navigator application, the DSA is displayed without subtraction. The unsubtracted images contain more anatomical information than the DSA, which may not be suitable for verification of the registration of a pre-operative CT/MR with x-ray images during the live guidance step.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime