Health Products

Giraffe Blue Spot PT Lite Phototherapy Systems (2019-12-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Datex-Ohmeda, Inc
9900 Innovation Drive
Wauwatosa
53226
Wisconsin
UNITED STATES
Giraffe Blue Spot PT Lite Phototherapy SystemsM1236303

Introduction

Affected Products

Giraffe Blue Spot PT Lite Phototherapy Systems

Reason

Some Giraffe Blue Spot PT Lite Phototherapy Systems built between 2014 and Jan 2018, have a light output falling below the recommended minimum output for the Giraffe Blue Spot PT. This problem has been associated with exposure of the light pipe to elevated temperatures, which is contra-indicated in the operation, maintenance, and service manual. There have not been any reported harms or adverse events associated with the Giraffe Blue Spot PT Lite.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime