MultiDiagnost Eleva- Flat Detector-Image Detector System - 2019-12-27

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
F. Allura Xper FD10/10(Product of Alluraclarity Family)722027
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
D. Allura Xper FD10(Product of Alluraclarity Family)722026
Not applicable. 
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
E. Allura Xper FD20(Product of Alluraclarity Family)722028
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
B. OmniDiagnost Eleva708028
Not applicable. 
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
C. Allura Xper FD101443
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
A. MultiDiagnost Eleva- Flat Detector-Image Detector0708MD02

Introduction

Affected Products
  1. MultiDiagnost Eleva- Flat Detector-Image Detector
  2. OmniDiagnost Eleva
  3. Allura Xper FD10
  4. Allura Xper FD10(Product of Alluraclarity Family)
  5. Allura Xper FD20(Product of Alluraclarity Family)
  6. Allura Xper FD10/10(Product of Alluraclarity Family)

Reason

Philips identified that a capacitor inside the convertor of the velara x-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible anymore. The failed capacitor may produce smoke and a burning odor in the room where the generator is located. If the system is a biplane model, x-ray acquisition is still possible on the other channel. If the system is a monoplane, it cannot be used anymore until the capacitor is replaced.