F. Allura Xper FD10/10(Product of Alluraclarity Family)
722027
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
D. Allura Xper FD10(Product of Alluraclarity Family)
722026
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
E. Allura Xper FD20(Product of Alluraclarity Family)
722028
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
B. OmniDiagnost Eleva
708028
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
C. Allura Xper FD10
1443
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
A. MultiDiagnost Eleva- Flat Detector-Image Detector
0708MD02
Introduction
Affected Products
MultiDiagnost Eleva- Flat Detector-Image Detector
OmniDiagnost Eleva
Allura Xper FD10
Allura Xper FD10(Product of Alluraclarity Family)
Allura Xper FD20(Product of Alluraclarity Family)
Allura Xper FD10/10(Product of Alluraclarity Family)
Reason
Philips identified that a capacitor inside the convertor of the velara x-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible anymore. The failed capacitor may produce smoke and a burning odor in the room where the generator is located. If the system is a biplane model, x-ray acquisition is still possible on the other channel. If the system is a monoplane, it cannot be used anymore until the capacitor is replaced.