Health Products

Getinge 88-Series Washer/Disinfector (2020-01-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Getinge Disinfection AB
Ljungadalsgatan 11
Växjö
35115
SWEDEN
Getinge 88-Series Washer/Disinfector88-5

Introduction

Affected Products

Getinge 88-Series Washer/Disinfector

Reason

The washer disinfectors affected by this field action can develop cracks in spot welds inside the chamber. The risk for users is due to localization infection from contaminated water (from pre-rinse) ending up on the floor although of from slipping. Although this risk is already considered mitigated with provided information in the IFU on requirements.