Health Products

Rusch Greenlite (2020-01-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1903341
Manufacturer
Teleflex Medical
IDA Business and Technology Park, Dublin Road, Athlone, Co.
Westmeath
IRELAND
Rusch Greenlite4551004

Introduction

Affected Products

Rusch Greenlite

Reason

Teleflex has received complaints that the lights guide of the Rusch Greenlite Mac Laryngoscope broke at the point of a welded joint either prior to use during setup or during use inside the patients mouth.