Epidural Catheterization Kit (2020-01-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.B. Epidural Catheterization Kit
Manufacturer
Arrow International Inc.
2400 Bernville Road
Reading
19605
Pennsylvania
UNITED STATES
AJ-05500
13F18J0916C. Loss of Resistance Syringe
Manufacturer
Arrow International Inc.
2400 Bernville Road
Reading
19605
Pennsylvania
UNITED STATES
LR-05501
More than 10 numbers, contact manufacturer.A. Epidural Catheterization Set
Manufacturer
Arrow International Inc.
2400 Bernville Road
Reading
19605
Pennsylvania
UNITED STATES
ASK-05400-CA1

Introduction

Affected Products
  1. Epidural Catheterization Set
  2. Epidural Catheterization Kit
  3. Loss of Resistance Syringe

Reason

Arrow international, a subsidiary of Teleflex, is recalling the devices due to receipt of complaints reporting various failures with the LOR (loss of resistance) syringes resulting in dural puncture with some also requiring a blood patch.

Failures may present as follows:
  • the syringe may leak which could lead to a false positive result,
  • i.e. the clinician will think they have reached the epidural space before they do reach it.


The syringe may be blocked/sticking which may lead to a false negative result, as it would be possible to enter the epidural space with the epidural needle and be unaware of having done so because there would be no loss of resistance. In the latter setting, it would be possible to enter the subarachnoid space unknowingly, risking injury to nerves. This potential harm is also highly dependent on the level at which the procedure is performed. When performed for the relief of pain during labor and delivery, the procedure is ordinarily performed below the level of termination of the spinal cord, whereas if performed for the purpose of providing epidural analgesia following thoracic surgery for example, there is the risk of spinal cord injury.