Arrow international, a subsidiary of Teleflex, is recalling the devices due to receipt of complaints reporting various failures with the LOR (loss of resistance) syringes resulting in dural puncture with some also requiring a blood patch.
Failures may present as follows:
-
the syringe may leak which could lead to a false positive result,
-
i.e. the clinician will think they have reached the epidural space before they do reach it.
The syringe may be blocked/sticking which may lead to a false negative result, as it would be possible to enter the epidural space with the epidural needle and be unaware of having done so because there would be no loss of resistance. In the latter setting, it would be possible to enter the subarachnoid space unknowingly, risking injury to nerves. This potential harm is also highly dependent on the level at which the procedure is performed. When performed for the relief of pain during labor and delivery, the procedure is ordinarily performed below the level of termination of the spinal cord, whereas if performed for the purpose of providing epidural analgesia following thoracic surgery for example, there is the risk of spinal cord injury.