Atellica CH Reaction Cuvette Segment (2020-02-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
N1503218
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
Atellica CH Reaction Cuvette Segment11099326

Introduction

Affected Products

Atellica CH Reaction Cuvette Segment

Reason

Siemens Healthcare Diagnostics has determined that a small percentage (<0.5%) of Atellica CH Reaction Cuvette Segment Kit lots ending in "17" or "18" may have cuvette defects allowing water from the water bath to contaminate the interior of the cuvette. Not all cuvette positions within an affected cuvette segment are impacted. There is a potential that sample results obtained at the impacted cuvette positions can be falsely elevated or depressed to varying degrees depending on the assay and the amount of water bath contamination. Quality control samples may or may not detect this issue.