Health Products

Ysio Radiography X-Ray System - Full Motorized (2020-02-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Laceration Hazard
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Plus de 10 numéros, contactez le fabricant.
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
Ysio Radiography X-Ray System - Full Motorized10281013

Introduction

Affected Products

Ysio Radiography X-Ray System - Full Motorized

Reason

Siemens is notifying users of Ysio Fully Automated System with software version VB10 about a potential malfunction of the motor controller.
In rare cases, the system processes incorrect internal values for the position of its ceiling tube stand. If in this situation the user then activates a longitudinal or transversal stand movement (by using the buttons at the tube head or the remote console), the system's controls may (depending on the values being processed) initiate a ""correction"" of the tube stand position, resulting in an unexpected movement.