Health Products

LIAISON MDX/Integrated Cycler (2020-02-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lots.LIAISON MDX/Integrated Cycler
Manufacturer
Diasorin Inc.
1951 Northwestern Avenue, P.O. Box 285
Stillwater
55082
Minnesota
UNITED STATES
MOL1001

Introduction

Affected Products

LIAISON MDX/Integrated Cycler

Reason

Magnetic field produced by the magnets (Platen and Consumable Cover) in the LIAISON MDX/Integrated Cycler may exceed 0.5 mT (5 Gauss). Workers with implanted ferromagnetic or electronic medical devices should not be exposed to static magnetic fields exceeding 0.5 mT (5 Gauss).