Expression MR400 MRI Patient Monitoring System (2020-02-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.Expression MR400 MRI Patient Monitoring System
Manufacturer
Invivo, a Div. of Philips Medical Systems
12151 Research Parkway
Orlando
32826
Florida
UNITED STATES
866185

Introduction

Affected Products

Expression MR400 MRI Patient Monitoring System

Reason

For MR400 devices with software revisions 01.03.00, 01.04.00, 01.05.00 And 01.07.00, the menu selections for users to access the O2 sensor calibration were changed. These changes were not incorporated into the device's instructions for use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with these software revisions.