Health Products

XN-10 & XN-11 Modules (2020-02-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Sysmex Corporation
1-5-1 Wakinohama-kaigandori
Chuo-ku
651-0073
JAPAN
B. XN-11 ModuleXN-11
More than 10 numbers, contact manufacturer.
Manufacturer
Sysmex Corporation
1-5-1 Wakinohama-kaigandori
Chuo-ku
651-0073
JAPAN
A. XN-10 ModuleXN-10

Introduction

Affected Products
  1. XN-10 Module
  2. XN-11 Module

Reason

When the result database reaches 100,000 samples, the system may unexpectedly stop and generate an error message. Only software version 22.04-10 Is affected.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime