Health Products

ITS4850A Intellivue Telemetry System (2020-02-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
ITS4850A Intellivue Telemetry SystemTRX4851A

Introduction

Affected Products

ITS4850A Intellivue Telemetry System

Reason

The ECG signal from patients being monitored using a Philips TRx4851A Telemetry Transceiver may not be properly analyzed when the transceiver is used with a Philips Patient Information Center iX Release C.02.00, C.02.02, C.02.03 (all released versions of C.02, collectively referred to as ‘C.02.xx’) or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms.

Alarming based on pulse oximetry (SpO2) signals are unaffected by this issue.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime