PrisMax Control Unit (2020-03-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Baxter Healthcare SA
PO BOX 8010
Zurich
8010
SWITZERLAND
PrisMax Control Unit955558

Introduction

Affected Products

PrisMax Control Unit

Reason

Manufacturer received questions from clinicians exploring modifying their use of Prismax Control Units, in order to minimize exposure to COVID-19-positive patients. For example, clinicians may be using multiple extension lines to extend the length of the tubing set to allow placement of a Prismax Control Init outside of the patient's room. As per manufacturer, several significant risks could arise with this practice. To mitigate these risks, manufacturer is recommending following the setup instructions in the graphical user interface, as well as the warnings from the prismax operators  manual. Failure to do so could result in serious patient harm.