Health Products

ARCHITECT iGentamicin Reagent Kit (2020-04-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Abbott GMBH & Co. K.G.
MAX PLANCK RING 2
WIESBADEN
65205
GERMANY
A. ARCHITECT iGentamicin Reagent Kit1P31-25

Introduction

Affecetd Products
  1. ARCHITECT iGentamicin Reagent Kit

Reason

This recall is to inform customers of a product correction for the architect iGentamicin assay, and to provide instructions on the actions the laboratories must take.

Abbott identified that samples tested for architect stat high sensitive troponin-I and stat high sensitivity troponin-I assays (LN 3P25 and LN 2R98) may show interaction when processed directly after the architect iGentamicin (LN 1P31) assay and patient results might be impacted. Preliminary results from Abbott internal analysis of field data from these two assays show false elevation of patient results with a frequency of 0.002%. 

At this time, there are no indications that the issue impacts other assays, however, the investigation is ongoing. Further, corrective actions may be implemented and communicated upon completion of the investigation.
 