Health Products

Artis Pheno System (2020-05-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
164101
Manufacturer
Siemens Healthcare GMBH
Henkestr. 127
Erlangen
91052
GERMANY
Artis Pheno SystemNot applicable.

Introduction

Affected Products

Artis Pheno System

Reason

In the event of a fault within the drivetrain of an axis, the potential exists for the C-arm to leave its intended travel path by up to 10cm. The problem has not been reported at customer sites and is currently classified as a single event.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime