B. ADVIA Chemistry System Fructosamine (FRUC) Reagent
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
10361941
More than 10 numbers, contact manufacturer.
A. ADVIA Chemistry System Enzymatic Creatinine_2 (ECRE_2) Assay
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
4992596
Introduction
Affected Products
A. ADVIA Chemistry System Enzymatic Creatinine_2 (ECRE_2) Assay
B. ADVIA Chemistry System Fructosamine (FRUC) Reagent
Reason
Siemens healthcare diagnostics has become aware that the presence of n-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples. This interference may cause falsely depressed fructosamine and creatinine results in patient samples.
Currently, the "limitations of the procedure" sections in the affected assay instructions for use (IFU) do not indicated that NAPQI is a metabolite of acetaminophen and that certain concentrations of NAPQI (NAPQI concentrations specific to each affected assay) may lead to falsely depressed results for patient samples. The "limitations of the procedure" sections also do not state that the use of the assay is not recommended for patients being treated for an overdose of acetaminophen.